| Technical Specifications | TECHNICAL SPECIFICATIONS - SPIROMETRY Power supply: Rechargeable battery, approx. 10 hours autonomy Battery charger: 100-240 V, 50-60 Hz, 400-200 mA Temperature sensor: semiconductor (0-45°C) Flow sensor: bi-directional digital turbine Flow range: ±16L/s ±5% or 200mL/s Volume accuracy: ±2.5% or 50mL Dynamic resistance: <0.5 cm H2O/L/s Connectivity: USB 2.0, Bluetooth® Display: 7” colour touch screen LCD, 800x480 resolution Mouthpieces: Ø 30 mm (1.18 inch) Electrical protection: Class II Norms: Electrical safety standard EN 60601-1 Electromagnetic compatibility EN 60601-1-2
SPIROMETRY PARAMETERS FVC, FEV1, FEV1/FVC, FEV1/VC, PEF, FEF25, FEF50, FEF75, FEF25-75, FEF75-85, Lung Age, Extrapolated Volume, FET, Time to PEF, FEV0.5, FEV0.5/FVC, FEV0.75, FEV0.75/FVC, FEV2, FEV2/FVC, FEV3, FEV3/FVC, FEV6, FEV1/ FEV6, FEV1/PEF, FEV1/FEV0.5, FIVC, FIV1, FIV1/FIVC, PIF, FIF25, FIF50, FIF75, FEF50/FIF50, VC, IVC, IC, ERV, IRV, Rf, VE, VT, tI, tE, VT/tI, tE/tTOT, MVV (measured), MVV (calculated).
TECHNICAL SPECIFICATIONS - OXIMETRY SpO2 range: 0-99% SpO2 accuracy: ± 2% between 70-99% SpO2 Pulse Rate range: 18-300 bpm Pulse Rate accuracy: ± 2 BPM or 2% whichever is greater Memory capacity: up to 500 hours oximetry
OXIMETRY PARAMETERS %SpO2 and Pulse Rate (Min, Max, Average), Test duration, Total SpO2 Events, T90% (SpO2 time =89%), T89% (SpO2 time =88%), T40 (Bradycardia duration with Pulse Rate <40 BPM), T120 (Tachycardia duration with Pulse Rate >120 BPM) |
|