| Technical Specifications | ECG Leads: 3-6-12 channel illustration of ECG signals in continuous recording or 10 sec. recording with interpretation. Dimension: 200x300x97.5 mm Approx 3 kg Recording channel: 3, 6, 12 channel and 1 channel (60 sec.) Sensitivity: 5, 10, 20, auto (I~aVF: 10, W1~V6: 5) mm/mV Printing speed: 12,5; 25; 50 mm/s Filters: AC (50/60 Hz, -20 dB or better), Muscle (25~35 Hz, -3 dB or better) Base line drift (off 0.05, 0.1, 0.2 Hz - 3 dB or better) Low pass filter (off, 40 Hz, 100 Hz, 150 Hz) Sampling rate: analysis 500 samples/sec, digital 8,000 samples/sec Display: 2x16 character LCD display LCD display: Sensitivity, speed filter status, HR, Printing form, rhythm lead LED indicator: Signal quality, power source, battery Keyboard: Membrane keyboard, alphanumeric and symbol available Patient data: ID, name, age, sex, height, weight Basic measurement: Heart rate, PR, QRS, QT/QTc, P-R-T axis Recorder: Thermal print head, thermal paper roll Report papers: A4: 215X297 mm (8.5”x11.7”) Letter: 215x280 mm (8.5”x11”) Resolution: vertical: 8 dots/mm, horizontal: 16 dots/mm Electrical: Internal noise: 20 µV (p-p) max input impedance: >50MO Input voltage range: ± 5 mV CMRR: greater than 105 dB DC offset voltage: = ± 400mV Patient leakage current: < 10µA Frequency response 0,05~150 Hz within -3 dB Isolated, defibrillation and ESU protected Power supply: AC or rechargeable lithium-ion 11.1 V 2600 mA battery (optional) Power requirement: 100-240 Vac, 50/60 Hz, 1.0-0.5 A Power consumption: 60 W max Battery capacity: 6 hour of normal use (approx. 200 automatic ECG printouts) Communication: LAN Safety: Class I, Type CF Conformity: CE, FDA, KFDA, CFDA, NCC Made in Korea |
|